| Pumping type: |
Peristaltic finger pump |
| IV set requirement: |
Standard IV set (15,20 drops/ml), or micro drip IV set (60 drops/ml) with 3.4--4.5mm tubing |
| Flow rate setting range: |
1 ml/h--999 ml/h |
| Drip rate setting range: |
1 drop/min--333 drops/min |
| Flow rate step: |
0.1 ml/h |
| Flow rate accuracy: |
¡À5% ( standard PVC set must first be adjusted) |
| Drip rate accuracy: |
¡À1.5 % |
| Timing range: |
¡À30% |
| Volume setting range: |
0-9999ml |
| Volume infusion display range: |
0-9999ml |
| Purge rate: |
800 ml/h |
| KVO (keep vein open): |
1 ml/h |
| Air bubble detection method: |
Ultrasonic |
| Drip inspection method: |
Infrared photo-electric beam |
| Peristaltic finger pump equipment: |
Detachable |
| Display: |
Color LCD with high backlight |
| Block pressure detector: |
40KPa--140KPa, sensitive adjustment of 8 levels |
| Alarms: |
Infusion Monitoring: |
| Air in tube, door open, occlusion, pause, infusion finishing, abnormal dripping,leakage, tube slope etc. |
| Diagnosis malfunctions: |
| Low battery,damaging battery,abnormal dropping sense organ,abnormal electrical set. |
| AC power source: |
90V-240V, 50Hz/60Hz |
| Inner battery: |
Rechargeable Ni-hydrogen battery, DC12V, 2000mAh |
| Charge time: |
16 hours |
| Discharge time: |
¡Ý5 hours |
| Power consumption: |
AC power supply: <20 VA, battery supply: <5 W |
| Weight: |
2.1 kg |
| Safe classification: |
Class ; type CF; Portable; IPX 4 |
| Dimensions: |
140mm¡Á168mm¡Á230mm |
| Operational condition: |
Temperature: +5 -- +40; Relative humidity:20-90% |
| User range: |
Venous infusion,Vasodilator medicianl preparation,Anticancer drug,Accelerative procreation drug, Nutrient liquid and other kinds of medicinal liquid , mini infusion and rapid infusion. |